Surface treatment on the shear bond strength of dental ceramics to titanium commercially pure


1 Department of Dentistry, School of Dentistry, Federal University of Jequitinhonha and Mucuri Valleys, Diamantina, Brazil
2 Department of Prosthodontics and Dental Materials, School of Dentistry of Ribeirão Preto, University of São Paulo, São Paulo, Brazil

Date of Web Publication17-May-2017

Correspondence Address:
Maria da Gloria Chiarello de Mattos
Department of Dental Materials and Prosthodontics, School of Dentistry of Ribeirão Preto, University of São Paulo, Av. do Café s/n, Monte Alegre, 14040.904 Ribeirao Preto
Brazil
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2278-9626.206376

Rights and Permissions
  Abstract 

Background: The aim of this study was to evaluate ceramic/titanium shear bond strength and microstructure of two commercially dental ceramics. Materials and Methods: Ceramic/titanium bond strength (6.0 mm × 2.0 mm) was measured using a universal testing machine. The load was applied with a crosshead speed of 0.5 mm/min until fracture. The titanium received different modifications for commercially pure titanium (airborne-particle abrasion and acid and thermic treatment). Results: All specimens showed different features within the interfacial area according to the different kinds of surface treatment. The tensile bond strength showed significant difference among the ceramics and surface treatment where the acid treatment groups had least means bond strength as differences between ceramic groups (P = 0.03). Conclusion: It was concluded that the use of acid solutions did not improve titanium-ceramic bonding strength values and the combination of airborne-particle abrasion and thermic treatment produced the most significant improvement in the titanium-ceramic bonding.

Keywords: Bond strength, ceramics, commercially pure titanium, surface treatment


How to cite this article:
Galo R, Frizzas DG, Rodrigues RC, Ribeiro RF, de Mattos Md. Surface treatment on the shear bond strength of dental ceramics to titanium commercially pure. Eur J Gen Dent 2017;6:77-81

How to cite this URL:
Galo R, Frizzas DG, Rodrigues RC, Ribeiro RF, de Mattos Md. Surface treatment on the shear bond strength of dental ceramics to titanium commercially pure. Eur J Gen Dent [serial online] 2017 [cited 2017 Dec 18];6:77-81. Available from: http://www.ejgd.org/text.asp?2017/6/2/77/206376


  Introduction Top


Titanium commercially pure (CP-Ti) has desirable properties that would recommend in dental use when they are comparable with dental gold alloys.[1] Recently, the use of CP-Ti and its alloys in implants and prosthesis has increased. However, problems with ceramic bonding are encountered when CP-Ti is used in metal-ceramic restoration.[2] The oxidation characteristics of titanium are the main problems, especially because casting titanium surface is susceptible to oxygen contamination.[2] Special low-fusing dental ceramics are necessary for titanium-ceramic bonding.[3] Relatively thick and nonadherent layers of titanium oxide tend to form at temperature above 800°C;[4] therefore, ceramic should be fired below this temperature since a thick oxide layer apparently minimizes metal-ceramic bond strength.[5],[6] Other factors to be considered affecting titanium-ceramic bonding are adherence of the self-formed oxide to the titanium substrate; bonding by fusion of self-formed oxide to the ceramic; growth of an oxide layer on titanium at elevated temperatures;[1] and stress developed at the interface because of an imbalance in thermal coefficients that may affect the flexural bond strength of the titanium ceramic system.[7]

Several secondary factors may enhance the titanium-ceramic bond strength. These include alteration of the titanium surface using airborne-particle abrasion, acid etching,[8] and application of bonding agent before ceramic.[8],[9] Some studies show that titanium surface nitridation [10] or thin chromium coating,[11] bonding agents,[8] argon atmosphere during ceramic firing,[8],[12] and sputter-coated gold layer on titanium improved titanium-ceramic adhesion. The result is a highly activated surface that can be demonstrated by the wettability of the material.[13] Although airborne-particle abrasion of the titanium surface before ceramic application improves the adhesion of the ceramic to the titanium substrate,[9] this procedure remains technique sensitive. Dérand and Herø[4] observed that the use of larger alumina particles with a diameter of 250 μm, compared with 50 μm particles, significantly improved the bond between titanium and ceramic. It is possible that the small particles may embed in the titanium surface. In fact, Gilbert et al.[14] found that airborne-particle abrasion could contaminate the surface of titanium with alumina particles, which could weaken the mechanical interlocking of the ceramic and titanium. Contamination of the titanium surface might also decrease its corrosion resistance and biocompatibility.[15]

Galo et al.[16] evaluated the treatments based on airborne-particle abrasion (aluminum oxide [Al2O3]) of the titanium surface. The authors observed that specimens produced the highest mean repair bond strength values, dependent of the laser welded. Airborne-particle abrasion plays a critical role in composites bonding to CP-Ti by creating roughened surface and the combined between titanium and composite did increase the bond strength significantly when decreasing the aluminum particle. The objective of the present study was evaluating the effect of three different titanium surface treatments on the titanium-ceramic bond.


  Materials and Methods Top


Eighty specimens were cast with CP-Ti type I (Tritan; Dentaurum Inc., Pforzheim, Germany). Wax patterns were fabricated with dimensions of 6.0 mm × 2.0 mm, with cylindrical rods (2.0 mm diameter × 24.0 mm length) connected to the center of the disk.

Five patterns were sprued together and invested according to the manufacturer's instruction. Titanium casting was performed on the Rematitan Machine (Dentaurum, Pforzheim, Germany) by means of voltaic arc and injection through positive pressure of argon gas in the upper portion of the cylinder and negative through the vacuum in the lower cylinder. After casting, the CP-Ti specimens were divested and cleaned with carbide burs (702L; KG Sorensen Ind. Com. Ltda., Barueri, SP, Brazil) and airborne-Al2O3 abrasion was performed with particles of 110 μm in size for 20 s, under pressure of 30–40 psi, and distance of 3–5 cm.[16],[17],[18] Then, the specimens were divided into eight groups (n = 10) and subsequently submitted to one of the eight surface modification procedures [Table 1].
Table 1: Surface modifications for commercially pure titanium

Click here to view


Two low-fusing ceramics were used: Noritake Ti22 (Noritake Dental Supply Co Ltda., Nagoya, Japan) and Triceran (Dentaurum, Pforzheim, Germany). The specimens in each group were further divided into two subgroups of 5 each. Then, ceramic-bonding agent powder and liquid were mixed. A thin layer was painted with a short bristle brush on the specimen's surface. The bonding agent was dried at the muffle entrance of the furnace and fired according to the manufacturer's instruction.

The entire process followed the manufacturer's instructions and began with application of 2 uniform coats of opaque ceramic, using a brush on each treated ceramic-bearing surface. After the opaque ceramic firing cycle, the area that should receive the ceramic was demarcated and a 2.0 mm thick polyester spacer (Plexiglass; General Electric, Mt. Vernon) was used to obtain a standard dentin ceramic for each specimen. The firing shrinkage was compensated for by applying a second layer of body ceramic, yielding a final total thickness of 2.0 mm. A glazing procedure was not performed. In all specimens, a single investigator performed ceramic application.

Shear bond strength testing

For the shear bond test, a special stainless steel device was fabricated to enable the specimens to be held firmly during the shear bond test. The device containing the CP-Ti-ceramic specimen was placed in a universal testing machine (EMIC MEM 2000; EMIC Equipamentos e Sistemas de Ensaio Ltda, Sao Jose dos Pinhais, PR, Brazil). A perpendicular load as applied at a distance of 0.5 mm from the CP-Ti-ceramic interface using a crosshead speed of 0.5 mm/min.[19] The shear bond strength was calculated according to the following formula: T = F/S (MPa), where T = tensile strength; F = critical rupture load; and S = metal ceramic bond area.

Structure of dental ceramic

Specimens from each group were selected for examination of the CP-Ti-ceramic under scanning electron microscope (SEM) (PHILIPS/XL 30-FEG). These specimens were embedded in autopolymerizing acrylic resin. The specimens were sectioned in the midsection with a low-speed saw along their width so that the cross-sectional area could be examined. The sectioned specimens were ultrasonically cleaned for 10 min and manually polished with abrasive paper (Metprep, Coventry, UK) with grits up to 1000 followed by three polishing stages (1, 0.34, and 0.05 μm) with Al2O3 polishing paste (Arotec, São Paulo, Brazil). Before testing, specimens were immersed in Kroll's solution for 20 s. Then, the specimens were ultrasonically cleaned in ethanol for 10 min and in distilled water for 10 min. The samples were well dried with hot air and kept in the desiccator until the beginning of the test.

Another specimen from each group was selected for the examination of failed surfaces of the CP Ti-ceramic interface. These specimens were ultrasonically cleaned in distilled water during 10 min before the making of the SEMs.

Statistical analysis

Kruskal–Wallis and Mann–Whitney U-tests were used to compare results of the shear bond strength. All statistical analyses and calculations were at the 95% level of confidence using the SPSS statistical program (SPSS for Windows, SPSS Inc., Chicago, Ill, USA).


  Results Top


The means value and standard deviations of shear bond strength (MPa) for the experimental groups are presented in [Figure 1]. Statistical analyses showed that there were significant differences between the bond strengths of surface treatment and the ceramic groups (P = 0.03), and all the groups treated with acid had significantly lower bond strength than the groups treated without acid or based on the manufacturer's instructions.
Figure 1: The means value and standard deviations of shear bond strength (MPa) for the experimental groups

Click here to view


Significantly, difference results were observed considering the ceramic and surface treatment factors as well as their interactions. For the groups were received Noritake ceramic treated with 110 and 50 μm airborne-particle abrasion without acid treatment demonstrated the higher mean values, but no significant differences between Triceran groups. The acid treatment groups had least means bond strength with differences between ceramic groups, except the groups received 110 μm airborne-particle abrasion with thermic treatment.

In addition, there were significant differences between the bond strength on Triceran groups (P = 0.03). There was a significant difference only between the groups was received acid treatment, where demonstrated lower values compared to no acid treatment.

The increases of shear bond strength were observed on Noritake groups. The Noritake group showed a significant difference in shear bond strength with respect to surface treatment (P = 0.03). The highest mean bond strength value was found in the 60 mesh airborne-particle abrasions with thermic and without acid treatment. Moreover, there were no differences in the bond strength values with the following groups: 110 μm airborne-particle abrasions with and without thermic treatment and then 60 mesh airborne-particle abrasions without thermic and acid treatment. However, there were differences statically with the other groups.

All specimens showed different features within the interfacial area according to the different kinds of surface treatment applied. A photomicrograph of the interface-bonded specimens showed black's spaces in the groups that were received acid treatment [Figure 2], and the photomicrograph of tested specimens demonstrated that the fracture line was located more often between ceramic and metal (CP-Ti) for all the groups. All of the experimental groups exhibited small of ceramic retained on the surface. This observation indicated that the mode of bond failure was primarily adhesive between the ceramic and CP-Ti.
Figure 2: Scanning electron microscope showing the interface-bonded specimens with black's spaces in both groups (1 – Triceram and 2 – Noritake) that were received acid treatment (Treatment 3 – TA3; Treatment 4 – TA4; Treatment 7 – TA7; and Treatment 8 – TA8) (×200)

Click here to view



  Discussion Top


The use of surface airborne-particle abrasion and surface treatment enhanced the CP-Ti/ceramic bond. However, the assumption of better results for bond strength between CP-Ti and ceramic when acid treatments were used for the modification of the surface of the metal substrate was not confirmed. Paradoxically, the application of the manufacturer's protocol showed to be more effective in increasing the bond strength when compared to the all treatments. These observations demonstrated that the use of acid solution could cause a weakening of the bond strength mechanism when used on the titanium surfaces for 2 min. These results were not consistent with the findings from another study,[20] which acid treatment contributes to improve the CP-Ti/ceramic bond strength values. However, this result is similar with the other study,[3] where the acid treatment decreases the bond strength between CP-Ti/ceramic systems.

For the groups treated with thermic treatment only, the group treated with acid presented significantly lower bond strength values than the manufacturer's protocol and no acid treatment groups. The group treated with no acid treatment had lower values with no differences from group control. These observations demonstrated that the use of acid treatment could cause a weakening of the bond strength mechanism when used on the titanium surfaces. In addition, the use of thermic treatment did not improve Triceran ceramic adhesion to the titanium surface, when compared to the manufacturer's protocol.

The group treated by 110 μm airborne-particle abrasions did not produce significantly different on the shear bond strength than the 50 μm airborne-particle abrasion. Reyes et al. found that airborne-particle abrasion particles produced greater bond strength.[9] The study indicates that a modified titanium surface, when the surface receiving airborne-particle abrasion or thermic treatment, may lead to a satisfactory CP-Ti/ceramic bonding. This is in agreement with previous studies.[9],[14] Wang and Fung [11] have indicated that the surface treatment produces a strong, protective, and adherence oxide layer that is suitable for ceramic bonding.

Al Hussaini and Al Wazzan [21] observed that surface treatment with airborne-particle abrasion enhanced the bond strength of CP-Ti to ceramic and titanium surface treatment with acid produced no effect on the titanium/ceramic bond. In addition, Cai et al.[20] reported that combination of hydrofluoric acid with other treatment on CP-Ti surface increases the bond strength between ceramic and the metal substrate. However, these results are not consistent with the findings from these studies where the acid treatment of the surface improved decrease of CP-Ti/ceramic bonding.

Moreover, it is considered that such surface modifications changed the chemical natural of the CP-Ti surfaces and facilitated the adhesion of ceramic to the titanium through the chemical adhesion mechanism.[9]

SEM photomicrographs demonstrated the interface aspects of CP-Ti/ceramic before the rupture in all the groups. A photomicrograph of the interface-bonded specimens showed black's spaces in the groups that were received acid treatment [Figure 2], and the photomicrograph of tested specimens demonstrated that the fracture line was located more often between ceramic and metal (CP-Ti) for all the groups. These results are probably in agreement with Troia et al.,[3] who observed the use of acid solutions did not improve CP-Ti/ceramic bonding strength values when compared to the control group. It has been shown that weak area of the CP-Ti/ceramic bond was the excessive and nonadherent oxide layer at the interface, and this problem must be solved to obtain greater bond strength.[22] The present study results suggest that use of surface airborne-particle abrasion along with an appropriate bonding agent will provide the highest bond strength of the ceramic to the titanium tested.

Photomicrographs showed that acid treatment produced a greater degree of roughness in the surface of titanium than the control group [Figure 2]. Furthermore, the surface roughness may increase the stress concentration at the CP-Ti/ceramic interface and generate steep reentrant angles that may prevent complete wetting and result in voids at CP-Ti/ceramic interfaces.[23]

The present study results suggest that the use of surface airborne-particle abrasion along with an appropriate surface treatment will provide the highest bond strength of the ceramic to the titanium tested. A limitation of this study was that only two brands of low-fusing ceramic and one brand of titanium were tested; the findings related to these three products may not be extrapolated to similar materials. Furthermore, if measuring the oxide layer thickness had been part of the study, helpful observations might have been obtained to better understand the behavior of the materials. Therefore, the impossibility of previewing the behavior of restorations in the oral cavity creates the expectancy for other studies corroborating in the development of new techniques that raise bond strength values between ceramic and titanium.


  Conclusion Top


Within the limitations of this investigation and for the materials used in this study, the following conclusions were made:

  1. The use of acid solutions did not improve titanium-ceramic bonding strength values when compared to the control group that represented the manufacturer's protocol
  2. The Triceram ceramic bond strength decreased significantly with an applied thermic treatment in CP-Ti
  3. The combination of airborne-particle abrasion and thermic treatment produced the most significant improvement in the titanium-ceramic bond
  4. Titanium surface treatment with acid treatment, with or without bonding agent, produced no effect on the titanium-ceramic bond, as compared to the control specimens.


Acknowledgments

We would like to thank Fundação de Amparo a Pesquisa do Estado de São Paulo.

Financial support and sponsorship

We would like to thank FAPESP (No. 99/01618-9) for the financial support.

Conflcts of interest

There are no conflcts of interest.

 
  References Top

1.
Adachi M, Mackert JR Jr., Parry EE, Fairhurst CW. Oxide adherence and porcelain bonding to titanium and Ti-6Al-4V aloy. J Dent Res 1990;69:1230-5.  Back to cited text no. 1
    
2.
Könönen M, Kivilahti J. Fusing of dental ceramics to titanium. J Dent Res 2001;80:848-54.  Back to cited text no. 2
    
3.
Troia MG Jr., Henriques GE, Mesquita MF, Fragoso WS. The effect of surface modifications on titanium to enable titanium-porcelain bonding. Dent Mater 2008;24:28-33.  Back to cited text no. 3
    
4.
Dérand T, Herø H. Bond strength of porcelain on cast vs. wrought titanium. Scand J Dent Res 1992;100:184-8.  Back to cited text no. 4
    
5.
Pang IC, Gilbert Jl, Chai J, lautenschlager EP. Bonding characteristics of low-fusing porcelain bonded to pure titanium and paladium-copper aloy. J Prosthet Dent 1995;73:17-25.  Back to cited text no. 5
    
6.
Pröbster l, Maiwald U, Weber H. Three-point bending strength of ceramics fused to cast titanium. Eur J Oral Sci 1996;104:313-9.  Back to cited text no. 6
    
7.
low D, Sumii T, Swain M. Thermal expansion coefficient of titanium casting. J Oral Rehabil 2001;28:239-42.  Back to cited text no. 7
    
8.
Atsü S, Berksun S. Bond strength of three porcelains to two forms of titanium using two firing atmospheres. J Prosthet Dent 2000;84:567-74.  Back to cited text no. 8
    
9.
Reyes MJ, Oshida Y, Andres CJ, Barco T, Hovijitra S, Brown D. Titanium-porcelain system. Part III: Effects of surface modification on bond strengths. Biomed Mater Eng 2001;11:117-36.  Back to cited text no. 9
    
10.
Oshida Y, Hashem A. Titanium-porcelain system. Part I: Oxidation kinetics of nitrided pure titanium, simulated to porcelain firing process. Biomed Mater Eng 1993;3:185-98.  Back to cited text no. 10
    
11.
Wang RR, Fung KK. Oxidation behavior of surface-modified titanium for titanium-ceramic restorations. J Prosthet Dent 1997;77:423-34.  Back to cited text no. 11
    
12.
Sadeq A, Cai Z, Woody RD, Miler AW. Effects of interfacial variables on ceramic adherence to cast and machined commercialy pure titanium. J Prosthet Dent 2003;90:10-7.  Back to cited text no. 12
    
13.
Kern M, Thompson VP. Sandblasting and silica-coating of dental aloys: Volume loss, morphology and changes in the surface composition. Dent Mater 1993;9:151-61.  Back to cited text no. 13
    
14.
Gilbert Jl, Covey DA, lautenschlager EP. Bond characteristics of porcelain fused to miled titanium. Dent Mater 1994;10:134-40.  Back to cited text no. 14
    
15.
Miyakawa O, Watanabe K, Okawa S, Kanatani M, Nakano S, Kobayashi M. Surface contamination of titanium by abrading treatment. Dent Mater J 1996;15:11-21.  Back to cited text no. 15
    
16.
Galo R, Ribeiro RF, Rodrigues RC, Pagnano Vde O, de Mattos Mda G. Effect of laser welding on the titanium composite tensile bond strength. Braz Dent J 2009;20:403-9.  Back to cited text no. 16
    
17.
Galo R, Frizzas DG, Rodrigues RC, Ribeiro RF, de Mattos MG. Shear bond strength of dental ceramics to cast commercialy pure titanium. Braz J Oral Sci 2010;9:362-5.  Back to cited text no. 17
    
18.
Andrade Tarozzo lS, Chiarelo De Mattos Mda G, Faria Ribeiro R, Semprini M. Comparison of retentive systems for composites used as alternatives to porcelain in fixed partial dentures. J Prosthet Dent 2003;89:572-8.  Back to cited text no. 18
    
19.
Kim JT, Cho SA. The effects of laser etching on shear bond strength at the titanium ceramic interface. J Prosthet Dent 2009;101:101-6.  Back to cited text no. 19
    
20.
Cai Z, Bunce N, Nunn ME, Okabe T. Porcelain adherence to dental cast CP titanium: Effects of surface modifications. Biomaterials 2001;22:979-86.  Back to cited text no. 20
    
21.
Al Hussaini I, Al Wazzan KA. Effect of surface treatment on bond strength of low-fusing porcelain to commercialy pure titanium. J Prosthet Dent 2005;94:350-6.  Back to cited text no. 21
    
22.
Kimura H, Horng CJ, Okazaki M, Takahashi J. Oxidation effects on porcelain-titanium interface reactions and bond strength. Dent Mater J 1990;9:91-9.  Back to cited text no. 22
    
23.
Kely M, Asgar K, O'Brien WJ. Tensile strength determination of the interface between porcelain fused to gold. J Biomed Mater Res 1969;3:403-8.  Back to cited text no. 23
    


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
Conclusion
Introduction
Materials and Me...
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed1070    
    Printed19    
    Emailed0    
    PDF Downloaded203    
    Comments [Add]    

Recommend this journal