Energy consumption of Twisted File instrument used with rotary or reciprocating adaptive motion


Department of Endodontics, Faculty of Dentistry, Ondokuz May?s University, Samsun, Turkey

Date of Web Publication1-Apr-2016

Correspondence Address:
Taha Ozyurek
Department of Endodontics, Faculty of Dentistry, Ondokuz May?s University, Samsun
Turkey
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2278-9626.179537

Rights and Permissions
  Abstract 

Objective: The aim of this study was to compare the energy consumption of Twisted File (TF) instrument used with rotary or adaptive motion (AM). Materials and Methods: Forty S-shaped Endo Training Blocks (Dentsply Maillefer) were used. Twenty were prepared using TF 25.06 in rotary motion (RM) group, and 20 were prepared using TF Adaptive SM2 in AM group. While engine-driven endodontic motors were connected to a digital wattmeter, the required torque for root canal instrumentation was analyzed by evaluating the electrical power consumption of the endodontic engine. Electric power consumption (mW/h), elapsed time (s), and a number of pecking motions required to reach the full working length (WL) were calculated. The data was statistically analyzed using Mann-Whitney U-test (P < 0.05). Results: Electrical power consumption during the preparation was significantly different between the groups; RM group exerted less electric power than reciprocating AM group did (P < 0.001). The required time to reach the full WL was not statistically significant between groups (P = 0.137). Conclusion: Within the limitation of this study, RM group exerted less electric power than reciprocating AM group did.

Keywords: Electric power consumption, nickel-titanium, nickel-titanium rotary preparation, Twisted File Adaptive


How to cite this article:
Ozyurek T, Keskin C, Demiryurek EO. Energy consumption of Twisted File instrument used with rotary or reciprocating adaptive motion. Eur J Gen Dent 2016;5:65-8

How to cite this URL:
Ozyurek T, Keskin C, Demiryurek EO. Energy consumption of Twisted File instrument used with rotary or reciprocating adaptive motion. Eur J Gen Dent [serial online] 2016 [cited 2020 Aug 13];5:65-8. Available from: http://www.ejgd.org/text.asp?2016/5/2/65/179537


  Introduction Top


Nickel-titanium (NiTi) instruments have been accepted as a significant improvement in endodontics in terms of reducing shaping time, producing a more centered shaping, and minimizing iatrogenic errors. [1],[2],[3] However, intracanal separation of NiTi instruments unexpectedly is not an uncommon problem in clinical practice. [4] NiTi instruments usually show no visible signs of permanent deformation and instrument separation occur unexpectedly. [5] The lifespan of a NiTi instrument has been correlated with its operational stress accumulation. [6] Bending and torsional stresses may contribute to instrument fatigue, which leads to mechanical failure. [7],[8] Fatigues are usually begin with a microcrack formation originates from surface irregularities. These microcracks develop to produce inner defects, which will compromise the fracture strength of NiTi instruments. [9]

Manufacturers design new file systems in order to increase fracture resistance of NiTi instruments by changing the metallurgic properties of NiTi, manufacturing processes, making surface modifications, or selecting different kinematics during instrumentation. [10],[11],[12] NiTi endodontic instruments have been traditionally used with continuous rotation motion until the use of an F2 ProTaper file with reciprocation motion was proposed by Yared's study. [13] Reciprocating motion has been reported to extend cyclic fatigue resistance of NiTi instruments compared to continuous rotation motion. [14],[15] Twisted File Adaptive system (TF Adaptive) has been introduced as a new instrumentation technique, combining reciprocation and rotation motions, thereby utilizes both techniques during instrumentation (http://www.sybronendo.com/cms-filesystem-action?file=/sybronendo-pdf/TF_Brochure.pdf). Elements Motor (SybronEndo, Orange, CA) automatically detects when the instrument is stressed and changes its motion. When the instrument is not stressed, it performs interrupted continuous rotation, but the moment instrumentation stress is increased, motion of instrument changes to reciprocation with specially designed clockwise and counterclockwise angles. Reciprocating motion also helps to prevent instruments to lock and/or thread (screw) into the canal. Screw effect, which happens when the instrument tip binds and shaft continues to rotate, causes high levels of stress on the instrument, thereby increasing the risk of separation.

The electric power consumption of an endodontic motor has been correlated with instrument stress state during instrumentation. The manufacturer claims that TF Adaptive system load less stress on the instrument. Therefore, the aim of this study was to compare the energy consumption of TF endodontic instrument used with rotary or adaptive motion (AM). The null hypothesis was that there would be no difference between rotation and AMs regarding electric power consumption.


  Materials and methods Top


Forty standardized ISO 15, 0.02 taper, S-shaped Endo Training Blocks (Dentsply Maillefer) with 16-mm working length (WL) were used in this study. One expert operator, previously calibrated for pecking speed and pressure on the handpiece, performed all the instrumentation phases of this study. The Endo Training Block canals were scouted up to the WL with a #10 stainless steel K-File (Dentsply Maillefer) and randomly assigned to 1 of 2 groups:

  • The TF rotary motion (RM) group (n = 20): After performing glide path with a #15 stainless steel K-File (Dentsply Maillefer), the 25 mm TF 25.06 was used to prepare the simulated canal until the WL. Each instrument was used with the Elements Motor; the speed of rotation was 500 rpm and torque was 4 Ncm
  • The TF AM group (n = 20): After the performing glide path with a #15 stainless steel K-File, the 25 mm SM2 (25.06) was used to prepare the simulated canal until the WL. Each instrument was used with the TF Adaptive program of the Elements Motor.


In all samples, irrigation was performed with distilled water and a new instrument was used to prepare each simulated canal.

The Elements Motor was connected to a digital wattmeter (Energy-measuring tool; Rev Ritter GmbH, Frankfurt, Germany) and an electronic schedule in order to evaluate the electric power consumption required to reach the full WL. The electronic schedule was designed to quantify and subtract the electrical and mechanical power disturbances. The electrical power consumption (mW/h), number of pecking motions, and time (s) required to reach the full WL with a TF 25.06 or SM2 file were calculated for every specimen belonging to the 2 groups (RM and AM). The Shapiro-Wilk test showed that data were not distributed normally thus differences between groups were statistically analyzed using the Mann-Whitney U-test. The significance level was set at P = 0.05. All statistical analyses were performed using the SPSS for Windows 21.0 software package (SPSS, Inc., Chicago, IL, USA).


  Results Top


In an experimental procedure, any TF 25.06 or SM2 file was not damaged or separated in both groups. In both groups, pecking motion required to reach apex was same (n = 5). Electrical power consumption during the preparation was significantly different between the groups; RM group exerted less electric power than reciprocating AM group (P < 0.001). The Elements Motor consumed 5.09 mW/h (standard deviation = 0.45) and 6.56 mW/h (standard deviation = 0.4) in the RM and AM groups, respectively [Table 1]. RM group took less time (48.34 ± 17.07) to reach the full WL than AM group (51.71 ± 9.28), but this was not statistically significant (P = 0.137). A ledge formation occurred in one sample in RM group.
Table 1: Electric power consumption (mW/h), time (s), and number of pecking motions required to reach the full working length with rotary motion or adaptive motion

Click here to view



  Discussion Top


Literature review has shown that fracture incidence for stainless steel files ranges from 2% to 6%. [16],[17] In 1980, NiTi instruments were invented to facilitate root canal preparation and thereby to reduce procedural errors such as apical transportation and file separation. NiTi instruments are 2-3 times more resistant to elastic and torsional fracture than stainless steel files. [18] However, fracture incidence of instruments increased, as NiTi rotary instruments were developed. [4] By modifying NiTi files' mechanical features such as diameter, cross section, and tip, fracture incidence for NiTi files was sought to be decreased and mechanical features were tired to be improved. For the same purpose, mechanical features of NiTi alloys were modified and new alloys such as M-Wire were obtained. Furthermore, movement kinematics was considered to decrease fracture incidence of NiTi files. [19] For over a decade, NiTi instruments had been used in continuous RM when Yared [20] proposed the use of NiTi instruments in reciprocating motion. Studies have demonstrated that reciprocating motion increases cyclic fatigue resistance of NiTi files, compared to continuous RM. [14],[21] Recently released TF instruments operating with Elements Motor combine both reciprocating motion and continuous RM. The manufacturer claims that this "adaptive motion" minimalizes the stress on the file and allows it to perform root canal preparation more easily and safely. [22]

In this study, we calculated the energy consumed by both TF NiTi files in continuous RM and in AM to reach WL in S-shaped resin blocks. Since resin blocks cannot simulate the root dentin, studies performed on these blocks do not reflect the clinical condition. However, it is a frequently used method, as the use of simulated canals in resin blocks allows standardization and replication of studies. [17],[23],[24] Another purpose of the use of S-shaped resin blocks is to enable Elements Motor to perform reciprocating motion in AM mode by exposing TF files to stress in the canal.

The manufacturing company proposes the use of SM1, SM2, and if necessary SM3 files at WL for TF adaptive system, respectively. It is also suggested that conventional TF files should be used in crown-down technique (http://www.sybronendo.com/cms-filesystem-action?file=/sybronendo-pdf/TF_Brochure.pdf). There is no difference between conventional TF files and TF files using adaptive system in terms of manufacturing. Both files were manufactured by being twisted in R-phase. In our study, in order to obtain standardization, resin blocks were shaped by using TF 25.06 and SM2 files according to a single-file principle. During shaping procedures, ledge formation was observed in 1 resin block in RM group and it was replaced by a new sample. However, there were no procedural errors in AM group. Franco et al. [25] proved in their study that reciprocating motion offers better shaping ability compared to continuous RM.

The results of this study revealed that TF NiTi files used in continuous RM consume significantly less energy than those used in AM. Thus, the null hypothesis was rejected. On the other hand, no difference was found between two motion types in terms of time taken to reach WL and the times of pecking motion. There is no study comparing electric consumption of TF NiTi files in the literature. In a study, Berutti et al. [22] compared glide path formation capability of two different path file NiTi rotary files using an energy consumption calculation method. Following the comparison of two different path files, the researchers calculated the energy consumption for the same NiTi file during shaping procedure. They have claimed that root canal preparation procedure results in stress on files and the stress increases electric consumption. [22] The results of this study revealed that increase in electric consumption during shaping procedure performed by using AM may not be associated with the fact that AM in canal causes more stress on the file compared to continuous RM, since previous studies have reported that reciprocating motion increases the resistance of NiTi instruments to cyclic and torsional fatigue when compared to continuous RM. [14],[21],[26],[27]


  Conclusion Top


Within the limitation of this study, RM group exerted less electric power than reciprocating AM group did.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Glossen CR, Haller RH, Dove SB, del Rio CE. A comparison of root canal preparations using Ni-Ti hand, Ni-Ti engine-driven, and K-Flex endodontic instruments. J Endod 1995;21:146-51.  Back to cited text no. 1
    
2.
Short JA, Morgan LA, Baumgartner JC. A comparison of canal centering ability of four instrumentation techniques. J Endod 1997;23:503-7.  Back to cited text no. 2
    
3.
Walia HM, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of nitinol root canal files. J Endod 1988;14:346-51.  Back to cited text no. 3
    
4.
Parashos P, Messer HH. Rotary NiTi instrument fracture and its consequences. J Endod 2006;32:1031-43.  Back to cited text no. 4
    
5.
Shen Y, Cheung GS, Bian Z, Peng B. Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 2006;32:61-5.  Back to cited text no. 5
    
6.
Yared GM, Dagher FE, Machtou P, Kulkarni GK. Influence of rotational speed, torque and operator proficiency on failure of greater taper files. Int Endod J 2002;35:7-12.  Back to cited text no. 6
    
7.
Gambarini G. Cyclic fatigue of ProFile rotary instruments after prolonged clinical use. Int Endod J 2001;34:386-9.  Back to cited text no. 7
    
8.
Loizides AL, Kakavetsos VD, Tzanetakis GN, Kontakiotis EG, Eliades G. A comparative study of the effects of two nickel-titanium preparation techniques on root canal geometry assessed by microcomputed tomography. J Endod 2007;33:1455-9.  Back to cited text no. 8
    
9.
Kim HC, Cheung GS, Lee CJ, Kim BM, Park JK, Kang SI. Comparison of forces generated during root canal shaping and residual stresses of three nickel-titanium rotary files by using a three-dimensional finite-element analysis. J Endod 2008;34:743-7.  Back to cited text no. 9
    
10.
Zinelis S, Eliades T, Eliades G. A metallurgical characterization of ten endodontic Ni-Ti instruments: Assessing the clinical relevance of shape memory and superelastic properties of Ni-Ti endodontic instruments. Int Endod J 2010;43:125-34.  Back to cited text no. 10
    
11.
Gambarini G, Grande NM, Plotino G, Somma F, Garala M, De Luca M, et al. Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods. J Endod 2008;34:1003-5.  Back to cited text no. 11
    
12.
Kim HC, Yum J, Hur B, Cheung GS. Cyclic fatigue and fracture characteristics of ground and twisted nickel-titanium rotary files. J Endod 2010;36:147-52.  Back to cited text no. 12
    
13.
Yared G. Canal preparation using only one Ni-Ti rotary instrument: Preliminary observations. Int Endod J 2008;41:339-44.  Back to cited text no. 13
    
14.
De-Deus G, Moreira EJ, Lopes HP, Elias CN. Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J 2010;43:1063-8.  Back to cited text no. 14
    
15.
Kim HC, Kwak SW, Cheung GS, Ko DH, Chung SM, Lee W. Cyclic fatigue and torsional resistance of two new nickel-titanium instruments used in reciprocation motion: Reciproc versus WaveOne. J Endod 2012;38:541-4.  Back to cited text no. 15
    
16.
Gambill JM, Alder M, del Rio CE. Comparison of nickel-titanium and stainless steel hand-file instrumentation using computed tomography. J Endod 1996;22:369-75.  Back to cited text no. 16
    
17.
Hülsmann M, Peters OA, Dummer PM. Mechanical preparation of root canals: Shaping goals, techniques and means. Endod Topics 2005;10:30-76.  Back to cited text no. 17
    
18.
Kuhn G, Jordan L. Fatigue and mechanical properties of nickel-titanium endodontic instruments. J Endod 2002;28:716-20.  Back to cited text no. 18
    
19.
Gambarini G, Gergi R, Naaman A, Osta N, Al Sudani D. Cyclic fatigue analysis of twisted file rotary NiTi instruments used in reciprocating motion. Int Endod J 2012;45:802-6.  Back to cited text no. 19
    
20.
Yared G. In vitro study of the torsional properties of new and used ProFile nickel titanium rotary files. J Endod 2004;30:410-2.  Back to cited text no. 20
    
21.
You SY, Bae KS, Baek SH, Kum KY, Shon WJ, Lee W. Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals. J Endod 2010;36:1991-4.  Back to cited text no. 21
    
22.
Berutti E, Alovisi M, Pastorelli MA, Chiandussi G, Scotti N, Pasqualini D. Energy consumption of ProTaper next X1 after glide path with PathFiles and ProGlider. J Endod 2014;40:2015-8.  Back to cited text no. 22
    
23.
Berutti E, Cantatore G, Castellucci A, Chiandussi G, Pera F, Migliaretti G, et al. Use of nickel-titanium rotary PathFile to create the glide path: Comparison with manual preflaring in simulated root canals. J Endod 2009;35:408-12.  Back to cited text no. 23
    
24.
Kum KY, Spängberg L, Cha BY, Il-Young J, Seung-Jong L, Chan-Young L. Shaping ability of three ProFile rotary instrumentation techniques in simulated resin root canals. J Endod 2000;26:719-23.  Back to cited text no. 24
    
25.
Franco V, Fabiani C, Taschieri S, Malentacca A, Bortolin M, Del Fabbro M. Investigation on the shaping ability of nickel-titanium files when used with a reciprocating motion. J Endod 2011;37:1398-401.  Back to cited text no. 25
    
26.
Pedullà E, Grande NM, Plotino G, Gambarini G, Rapisarda E. Influence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments. J Endod 2013;39:258-61.  Back to cited text no. 26
    
27.
Gavini G, Caldeira CL, Akisue E, Candeiro GT, Kawakami DA. Resistance to flexural fatigue of Reciproc R25 files under continuous rotation and reciprocating movement. J Endod 2012;38:684-7.  Back to cited text no. 27
    



 
 
    Tables

  [Table 1]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Materials and me...
Results
Discussion
Conclusion
References
Article Tables

 Article Access Statistics
    Viewed2863    
    Printed69    
    Emailed0    
    PDF Downloaded353    
    Comments [Add]    

Recommend this journal